Gas Chromatographic Injectors
Introduction
The great analytical strength of capillary gas chromatography lies in its
high resolution.
Capillary columns have 1) more theoretical plates (a measure of column
resolving power or efficiency) per meter as compared to packed columns
and 2) since they have less resistance to flow they can be longer than
packed columns. This means that the average capillary column (30
meters long) has approximately 100,000 theoretical plates while the
average packed column (3
meters) has only 2500 plates.
But with this separation power comes some limitations: 1) Capillary
columns, because they have smaller diameters (0.05 to 0.53 mm) than packed
columns (2 to 4 mm), require relatively specialized injectors and ancillary
flow and pressure controllers and 2) capillary columns require a smaller
amount of sample than packed columns. While the average sample mass of
each component in a mixture that is separable by packed column GC can be
in the microgram range (10-6 grams) per injection, capillary
columns routinely only handle 50 nanograms (10-9 grams) of a
particular component or less, sometimes much less.
Overloaded Chromatography
This sample size requirement initially meant that if samples contained
components that were too concentrated for a capillary chromatographic analysis,
the sample had to be diluted before it was analyzed. Otherwise the column
would be overloaded by those high concentrated
components. An example of this appears in the first figure below. The clearly
overload peaks are indicated. And while some of the other components are
in the resolvable (not overloaded) range, having large masses of components
can also distort the peak shape of some of the lower mass components.
An Overloaded Chromatogram
The following figure shows a little better chromatography with fewer
overloaded peaks. The second eluting peak (about 6 minutes) is clearly
not overloaded while the group between 10 and 14 minutes still shows overloading
characteristics: long drawn-out tailing and much less than baseline separation
with peaks that elute nearby (the 11 and 12 minute peaks, for instance).
Normal Packed Column Injector
The normal sequence of events in a GC injection is as follows. We will
assume in this explanation that some analytes are dissolved in a (liquid)
solvent although much of this process also holds for gas GC injections
too: A small amount of liquid (microliters) is injected through a silicon
rubber septum (using a special microliter syringe) into the hot (usually
200+ degrees C) GC injector that is lined with an inert glass tube. The
injector is kept hot by a relatively large, metal heater block that is
thermostatically controlled. The sample is immediately vaporized and a
pressurized, inert, carrier gas-which is continually flowing from a gas
regulator through the injector and into the GC column-sweeps the gaseous
sample, solvent, analyte and all, onto the column. In the packed column
injector, ALL the vaporized sample enters onto the column. This is how
all packed column injectors work; everything that is injected goes onto
the column. One modification of this is a small ancillary flow of carrier
gas that bathes the underside of the injector's septum so that hot vaporized
sample gases can't interact and possibly stick to the septum. This improves
peak shape and reproducibility. This last feature is called the septum
purge. The following figure is a schematic of a packed column injector,
sometimes called a direct or flash injector. The septum purge is not shown
here although the carrier gas regulator and inlet, a septum,
and an injection port liner ARE detailed.
Last by not least, the packed GC column
itself is connected at the bottom of the injector via metal fittings.
Schematic of packed GC column injector with septum purge
One last subtle point about the configuration of the carrier gas inlet:
notice that it enters the injector at about the middle of the heating block
and its gas has to travel along the outside of the injector before it enters
the injection port liner AT THE TOP.
This is so the carrier is preheated before it enters the liner where the
sample is vaporized. This helps to prevent a cold spot at the top of the
injector where the carrier gas enters.
Now remember that the size of the capillary column limits the amount
of analyte that can be injected, otherwise, chromatographic overloading
occurs. Therefore this packed column injector design, if used with a capillary
column, would require that samples with high concentrations of analytes
be diluted. Unless... what other alternative is there to get the amount
of analytes that are injected onto the column smaller without having to
dilute concentrated samples? The solution is the split/splitless capillary
GC injector.
The Split/Splitless Capillary Injector
OK so far so good. But how does the split/splitless injector work? It starts
with the same requirements as the packed column injectors: carrier gas
inlet, a septum, septum purge, injector
insert, heater block, and column
connection; but the heart of this technological feat is another set of
gas lines out of the injector--another
path that the vaporized sample can take. This is called the split
line or vent. The manufacturers of
these systems design them so that the carrier gas flow onto the column
is constant--to maintain the chromatographic requirements of the column
and yield reproducible retention times for analytes. At the same time,
the amount of gas that goes out the split vent controls the amount of sample
that enters the column. If the split vent is closed, via a computer controlled
split
valve, then all of the sample introduced into the injector goes
on the column. If the split vent is
open then most of the vaporized sample is thrown away to waste via the
split vent and only a small portion of the sample is introduced to the
column.
The following diagram illustrates a split/splitless injector with the split
vent on so that only a small portion
of the sample injected goes on the column.
And finally, a very neat (and fundamental) aspect of this is that the
amount of gas exiting the split vent can be varied while keeping the flow
onto the column constant. This means that the AMOUNT of the split (called
the split ratio) can be varied (in modern instruments via software control).
A common split ratio is 50 to 1. That is, for every 50 units of gaseous
sample that are thrown away to waste, 1 unit goes on (in) the column.
The analyst keeps careful control of the split ratio so that results
from the chromatography can still be quantified. Using a split ratio of
50, chromatographic peaks that show
up as, say 2.5 ng of compound X (using a calibration curve of detector
response versus peak area) really represent 2.5 x 50 = 125 ng of analyte
X in the original sample (the split ratio
was 50 remember?). Also notice that this analyte mass (125 nanograms) would
have overloaded the column if all of it ended up on the capillary column.
Voila! A split injection, and no sample dilution required.
These notes were written by Dr.
Thomas G. Chasteen at Sam Houston State University, Huntsville, Texas.